FEATURES
● History records
● Unique intelligent self-recognition technology of syringes
● Large & colorful LCD display
● RS232 interface
● Adjustable buzzer volume
● Anti-bolus unction
● TIVA mode
● 90º rotatable pole clamp convenient for horizontal bar & vertical IV pole
● Dynamically display the pessure
● User self-define syringe brands
● Stackable function
SPECIFICATIONS
Infusion rate |
5 ml syringe: 0.1-150ml/h 10 ml syringe:0.1-300ml/h 20 ml syringe: 0.1-600ml/h 30 ml syringe; 0.1-900ml/h 50 ml syringe: 0.1-1500ml/h |
Syringe Accuracy |
±2%(after correct calibration) |
Applicable syringes |
5, 10, 20, 30, 50(60)ml various brand of sterile syringe |
Volume to be infused(VTBI) |
0~9999.9ml |
Volume infused (Σ) |
0~9999.9 ml |
KVO rate |
0.1-2ml/h adjustable;default value 0.1ml/h |
Bolus rate |
5ml syringe:100-150ml/h 10ml syringe:100-300ml/h 20ml syrnge;100-600ml/h 30ml syringe:10-900ml/h 50/60ml syringe:100-1500ml/h |
Infusion Mode |
Rate mode Dose mode Time mode Body weight mode Drug library mode Intermittent mode TIVA Optional |
Occlusion pressure |
(40.0160.0)Kpa |
Auible and visual alarm |
Syringe disengaged, Infusion completion, empty, occlusion, Near completion Low battery, handle voff,No AC power supply,Malfunction, batteryexhausted etc. |
Nurse call interface |
Optional |
Power supply |
AC: 100-240v, 50/60Hz Dc:12v ± 1.2V |
Battery |
Lithium polymer 7.4V 1900mAh,running about 6 hours at 5ml/h |
Power consumption |
25VA |
Safe classification |
Class 1, CF |
Waterproof level |
IP x 3 |
Dimension |
230*137*139(L*W*H) |
Net weight |
1.7kg |
Operating conditions |
Temperature: 5ºC-40ºC, Relative humidity 10-95%, Atmosphere pressure:86-106Kpa |
Transport Conditions |
Temperature: -20ºC-60ºC, Relative humidity 10-95%, Atmosphere pressure:86-106Kpa |
Storage conditions |
Temperature: -20ºC-45ºC, Relative humidity 10-95%, Atmosphere pressure:86-106Kpa |